FEP630, FEH630 ELECTROMAGNETIC FLOWMETER  |  OI/FEP630/FEH630-EN REV. D  21 
 
 
 
 
Installing the sensor 
 
 
NOTICE 
Damage to the device 
Damage to the device due to improper assembly. 
•  The use of graphite with the flange or process connection 
gaskets is prohibited. This is because, in some instances, 
an electrically conductive coating may form on the inside 
of the meter tube. 
•  Vacuum shocks in the piping should be avoided to prevent 
damage to the liners (PTFE liner). Vacuum shocks can 
destroy the device. 
 
The flowmeter sensor can be installed at any location in the 
piping while taking the installation conditions into account. 
1. Remove protective plates, if present, to the right and left of 
the meter tube. To prevent possible leakage, make sure that 
the liner on the flange is not cut or damaged. 
2. Position the flowmeter sensor plane parallel and centered 
between the piping.  
3. Install gaskets between the surfaces, see Gaskets on 
page 16. 
 
Note 
For achieve the best results, ensure the gaskets fit concentrically 
with the meter tube 
To ensure that the flow profile is not distorted, the gaskets must 
not protrude into the piping. 
 
 
4. Use the appropriate screws for the holes in accordance with 
Torque information on page 144. 
5. Slightly grease the threaded nuts. 
6. Tighten the nuts in a crosswise manner as shown in the 
figure. Observe the tightening torques in accordance with 
Torque information on page 144! 
  First tighten the nuts to approx. 50 % of the maximum 
torque, then to 80 %, and finally a third time to the maximum 
torque. Do not exceed the max. torque. 
 
G11726
1
2
7
8
5
3
4
6
1
2
3
4
 
Figure 23:  Tightening sequence for the flange screws