FEM630 | ELECTROMAGNETIC FLOWMETER  |  OI/FEM630-EN REV. A
9     ...Diagnosis / error messages
...Extended diagnostic functions
Monitoring the electrode impedance
The measurement monitors the impedance between the 
measuring electrode and grounding and activates an alarm 
if the impedance drops below a limit. The function is 
activated together with the conductivity measurement.
Requirements for use:
•  
cm.
Additional installation conditions:
•  When using plastic piping, install a grounding plate at the 
front and back of the device.
•  There must not be any deposits on the measuring elec-
trodes.
•  The measuring tube must always be completely full, and 
the measuring medium must have only minor deviations in 
conductivity.
Measurements on the flowmeter
Coil inductance, coil current, coil resistance
The diagnosis of the coil in the sensor includes coil 
inductance, current and resistance.
Flowmeter sensor temperature
The coil temperature monitoring triggers an alarm via the 
digital output, if configured.
The minimum and maximum alarm value for the coil 
temperature can be set.
The coil temperature is a function of the ambient 
temperature and measuring medium temperature.
Compliance with the temperature specification of the 
sensor liner can thus be monitored.
Setup
Transmitter monitoring
Monitoring the temperature of the electronic unit in the 
transmitter triggers an alarm via the digital output, if 
configured.
In the ‘...Diagnosis Values’, the current temperature as well 
as the smallest and largest previously measured 
temperature is displayed.
Menu / parameter
Diagnostics / ...Diagnosis Control / ...Diagnosis Coil
Coil Diag On/Off Activate the function.
Coil Resistor Display the coil resistance.
Coil Current Display the coil current.
Coil Inductance Display the coil inductance.
Coil Temperature
Display the coil temperature within the 
sensor.
Coil Temperature Adj
Measurement of coil temperature must be 
set in accordance with the conditions on-
site. Temperature measured with a separate 
thermometer can be entered here.
Coil Temp. Min Alarm
Min. and max. alarm for the sensor 
temperature (coil temperature). Can be used 
to monitor the temperature limit of the 
meter tube liner
Coil Temp. Max Alarm