Grounding
The flowmeter sensor must be connected to ground 
potential. For technical reasons, this potential must be 
identical to the potential of the measuring medium.
In piping made of plastic or with insulating liner, grounding of 
the measuring medium is done by installing grounding plates. 
If stray potential is present in the piping, adding a grounding 
plate on both ends of the flowmeter sensor is recommended.
Inlet and outlet sections
The measuring principle is independent of the flow profile as 
long as standing eddies do not extend into the measured 
value formation, such as may for example occur after double 
manifolds, in the event of tangential inflow, or where half-
open gate valves are located upstream of the sensor. In such 
cases, measures must be put in place to normalize the flow 
profile.
Double manifold
Turn-off device
1
2
Figure 12  Inlet and outlet section, turn-off devices
A  
B 
Do not install fittings, manifolds, valves, etc., right before 
the flowmeter sensor.
Inlet / outlet sections: length of the straight piping 
upstream and downstream on the sensor. 
Experience has shown that, in most installations, straight 
inlet sections 3 × DN long and straight outlet sections 2 × 
DN long are sufficient (DN = nominal diameter of the 
flowmeter sensor). 
For test stands, the reference conditions of 10 × DN 
straight inlet and 5 × DN straight outlet must be provided, 
in accordance with EN 29104 / ISO 9104. 
Valves or other turn-off devices should be installed in the 
outlet section. 
Valve flaps must be installed so that the valve damper 
plate does not extend into the flowmeter sensor.