54
FEM630 | ELECTROMAGNETIC FLOWMETER  |  OI/FEM630-EN REV. A
...7     Commissioning
Configuration via the infrared service port adapter on the 
service port adapter.
By combining the HART-DTM and the software ‘flow’ available 
at www.abb.com/ABB AssetVision, all parameters can also be 
set without a HART connection.
For more detailed information on operating the software and 
the HART modem, please refer to the relevant operating 
instructions and the DTM online help.
Configuration via the HART interface of the device requires a 
PC / Notebook and a suited HART® Modem.
All parameters can also be set via the HART protocol, using 
the HART DTM available at www.abb.com/flow and theABB 
AssetVisionsoftware.
1  Position the infrared service port adapter on the front 
plate of the transmitter as shown
2  Insert USB interface cable into a free USB female 
connector on the PC / notebook.
3  Switch on the device power supply.
4  Start ABB AssetVision and perform the parameterization 
of the equipment.
Detailed information on operating the software is available in 
the relevant operating instructions and the DTM online help.
Infrared service port adapter
USB-interface cable
PC / Notebook running ABB AssetVision and HART DTM
PC / Notebook running ABB AssetVision and HART DTM
HART modem
Power supply unit
1
2
3
1
2
3
Figure 65  Infrared service port adapter 
on the transmitter (example)