40 S4 D184B122U02
Submenu/
Parameter
Submenu/
Parameter
Submenu/
Parameter
Selections
Value Range
Comments
Warning register
After the Warning Register is accessed the
text "Warning register" is displayed together
with the Quantity of warnings set. Using
ENTER again the set warnings with their
Warning Numbers and descriptive text are
displayed. Using ENTER again the question
is displayed if the warning is to be reset or if
the Help text is to be displayed. Using STEP
it is possible to display all the set warnings
with their Warning Number and descriptive
text. Using ENTER the set errors can be
reset.
Quantity: 2
3 (set)
ENTER * Test Mode *
Reset: Enter
Help text: Step
2
Totalizer reset
Display of the number of power outages
since start-up
Mains interrupt
0
Submenu
Syst. Adjust
Start System Zero
Manual
Auto. adjustment
Zero value check if required (required for
matching older model flowmeter primaries).
Manual entry, e.g. after a converter ex-
change.
Automatic adjustment:Valve must be
closed, liquid in flowmeter primary must be
at absolute zero flow. Initiate the automatic
adjustment with ENTER. The limits for the
zero value are ± 10%. If the value is outside
of these limits, the adjustment is rejected.
ENTER 0.0 %
System zero
numeric entry
0.0 %
FSM4000 08/2003
view only Display of the software revision. 08/2003
identifies the issue date.B. 10 = Revision
level for the Standard and HART-Protocol
software.
D699G004U01 B.10