EasyManua.ls Logo

ABB SM500F - Page 142

ABB SM500F
190 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SM500F
Field mountable paperless recorder 8 Specification
140 IM/SM500F Rev. Z
Analog/Digital Inputs
General
Number of inputs
7 (1 as standard, up to 6 optional)
Input types
mA, mV, voltage, resistance, THC, 3-wire RTD, 4-wire RTD, volt-free digital, 24 V DC digital*
Thermocouple types
B, E, J, K, L, N, R, S, T
Resistance thermometer
PT100
Other linearizations
x, x
3
/2, x
5
/2, custom linearization
Digital filter
Programmable 0 to 60s
Display range
–99999 to +999999
Common mode noise rejection
>120 dB at 50/60 Hz with 300 imbalance resistance
Normal (series) mode noise rejection
>60 dB at 50/60 Hz
CJC rejection ratio
0.05 °C/°C
Sensor break protection
Programmable as upscale or downscale
Temperature stability
0.02 %/°C or 2 µV/°C
Long term drift
<0.2 % of reading or 20 µV annually
Input impedance
>10 M (millivolts inputs)
>10 M (voltage inputs)
44 (mA inputs)
Analog/Digital resolution
16 bit

Table of Contents

Other manuals for ABB SM500F