EasyManua.ls Logo

ABB SM500F - Power Supply Connections

ABB SM500F
190 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SM500F
Field mountable paperless recorder 2 Installation
20 IM/SM500F Rev. Z
2.6 Power Supply Connections
Compensating Cable
BS1843 ANSI MC 96.1 DIN 43714 BS4937 Part No.30
Thermocouple Type + Case + Case + Case + Case
Ni-Cr/Ni-Al (K) Brown Blue Red Yellow Red Yellow Red Green Green Green White Green *
Ni-Cr/Cu-Ni (E) Violet White Violet *
Nicrisil/Nisil (N) Orange Blue Orange Orange Red Orange Pink White Pink *
Pt/Pt-Rh (R and S) White Blue Green Black Red Green Red White White Orange White Orange *
Pt-Rh/Pt-Rh (B) Grey White Grey *
Cu/Cu-Ni (T) White Blue Blue Blue Red Blue Red Brown Brown Brown White Brown *
Fe/Con (J) Yellow Blue Black White Red Black Red Blue Blue Black White Black *
* Case Blue for intrinsically safe circuits
Fe/Con (DIN 43710)
DIN 43710
Blue/red Blue Blue
Table 2.1 Thermocouple Compensating Cable
Fig. 2.11 Power Supply Connections
Note. Power supply terminal screws must be tightened to a torque of 0.8 Nm (7 lbf.in).
90 V min. to 264 V max. 50/60 Hz
10 V min. to 36 V max. DC