SM500F
Field mountable paperless recorder 8 Specification
142 IM/SM500F Rev. Z
Relays
Number of relays
1 as standard, 2 optional, (1 Module)*
* When using dual relay option boards, high voltages above 120 V AC must not be connected alongside low voltage DC
supplies.
Type and rating
For dual relay boards, only the following permutations are permitted.
30 V DC / 30 V DC
120 V AC / 30 V DC
120 V AC / 120 V AC
240 V AC / 240 V AC
240 V AC / 120 V AC
AC supplies must be on the same phase.
2-Wire Transmitter Power Supply (optional)
Number
2 isolated supplies
Voltage
24 V DC nominal
Drive
22 mA (each supply)
Ethernet Module (optional)
Physical medium
10BaseT
Protocols
TCP/IP, FTP (server), HTTP, SMTP, Modbus TCP (Client + Server)
FTP server functions
Directory selection & listing
File upload/download
4 independently configurable users with full or read-only access
Web server functions
Operator screen monitoring/selection. Remote monitoring of recording channels, analog/digital signals, alarms,
totalizers and archiving
SMTP client compatibility
Compatible with MS Exchange versions up to and including MS Exchange 2003
Relay type Selectable NO/NC
Voltage 250 V AC 30 V DC
Current 5 A AC 5 A DC
Loading (non-inductive) 1250 VA 150 W