ABB  TB82PH ATEX Product Instruction Manual 
TB82PH-ATEX-EN-A  11 
Figure 12 
MEASURE 
CALIBRATE 
OUT/HOLD 
CONFIGURE 
SECURITY 
DISPLAY 
25°
 
exit to
MEASURE 
FAULT 
info 
 
SELECT 
ENTER 
 
          YES 
          NEXT
         NO 
        MENU
%
mV 
pH 
 
 
 
OUTPUT 
HELD 
SPIKE  FAULT 
M 
ABB 
TB82 
Hidden Key 
User Mode
MEASURE 
CALIBRATE 
OUT/HOLD 
CONFIGURE 
SECURITY 
DISPLAY 
25°
 
        MENU
 
pH 
 
 
M 
Figure 13 
MEASURE 
CALIBRATE 
OUT/HOLD 
CONFIGURE 
SECURITY 
DISPLAY 
Calibr 
exit to 
MEASURE 
 
 
SELECT 
 
 
        MENU
 
pH 
 
 
 
Figure 14 
OPERATING PROCEDURES 
 
Figure 12 shows the TB82PH keypad and display with all possible icons 
in the display activated.   For any given operation only the icons involved 
in the operation will be shown. 
 
Key Functions
 
Exit to MEASURE - Escapes back to the Measure Mode from all other 
modes or programming states of operation. 
FAULT Info - Accesses information on diagnostic or error conditions. 
SELECT - Selects the mode or programming state of operation shown 
in the secondary display region 
ENTER - Accepts the mode or programming state of operation shown in 
the secondary display region 
NEXT - Increments through a series of programming states 
YES - Affirms the action that is about to take place 
NO - Denies the action that is about to take place. 
MENU - Increments through the modes of operation 
  Increases numerals & moves through a series of parameters 
  Moves the flashing data entry value one space to the right  
  Decreases numeric values & moves through a series of parameters 
 
Figure 13 shows the TB82PH in the MEASURE mode of operation.  
Pressing the key at the far right, underneath the MENU icon, provides 
access to all other instrument functions.   
 
Figure 14 shows the display after the MENU key has been pressed 
once. The CALIBRATE menu can now be accessed by pressing the key 
underneath the SELECT icon.   If the MENU key is pressed again the 
display will change to the next menu function choice, OUT/HOLD.  
Pressing again will activate the CONFIGURE menu choice, and so on.  
 
There are six MENU functions:
 
MEASURE - Normal operational display 
CALIBRATE – Allows calibration of the pH (ORP) and temperature 
sensors, restoring calibration to the original factory values, tuning of 
the analog output and also allows the ability to view or change the 
sensor’s slope and offset.  
OUT/HOLD - Allows the output to be re-ranged, held or released 
(Analog version only) 
CONFIGURE - Sets up how the transmitter operates  
SECURITY - Sets password protection to the CONFIGURE, 
CALIBRATE and OUTPUT menus 
DISPLAY - Displays temperature in °F or °C, current output in mA, 
reference impedance, mV input, spike output status and software 
revision.  Sets what will be displayed under the secondary display 
when in the measure mode. 
 
 
 
 
 
Commissioning the Instrument 
 
To commission the TB82:
 
1.  Configure the instrument as required for the application. 
2.  Calibrate the sensor(s). 
3.  Set up parameters on the receiving devices (e.g. PLC, DCS recorder) to properly recognize the TB82 outputs.