ABB  TB82PH ATEX Product Instruction Manual 
TB82PH-ATEX-EN-A  19 
SPECIFICATIONS 
 
Property  Characteristic/Value 
Process Display Range 
pH 
ORP 
pIon 
Temperature 
 
-2.00 to +16.00 pH 
-1999 to +1999 mV 
-1999 to +1999 mV 
 0 to 140°C
Power Requirements 
Analog  
 
 
HART Versions 
 
 
PROFIBUS PA 
 
 
FOUNDATION Fieldbus (FF) 
 
 
13.5 to 55 VDC (13.5 to 42 VDC for agency certified applications). 
See Figure 27 below 
 
14 to 55 VDC (14 to 42 VDC for agency certified applications). 
See Figure 27 below 
 
9 to 32 VDC (9 to 24 VDC for agency certified applications) 
15 mA quiescent current 
 
9 to 32 VDC (9 to 24 VDC for agency certified applications) 
15 mA quiescent current 
 
Ambient Temperature Effect 
pH 
ORP 
pION 
Temperature 
 
±0.007 pH/°C @ 95% Relative Humidity 
±0.4 mV/°C @ 95% Relative Humidity 
±0.4 mV/°C @ 95% Relative Humidity 
±0.16°C/°C @ 95% Relative Humidity 
Maximum Sensor Cable Length  100 ft (30.5 m) 
Environmental 
Operating temperature 
LCD Range 
Storage temperature 
 
-20° to 60°C (-4° to 140°F) 
-20° to 60°C (-4° to 140°F) 
-40° to 70°C (-40° to 158°F) 
Enclosure Classification  NEMA 4X 
IP65 
Size 
Height 
 
Minimum panel depth 
Maximum panel cutout 
Recommended panel cutout 
 
144 mm high x 144 mm wide x 171 mm long (5.67 in. high x 5.67 in. wide x 6.75 in. 
long) 
145 mm (5.70 in.) 
136.7 mm x 136.7 mm (5.38 in. x 5.38 in.) 
135 mm x 135 mm (5.33 in. x 5.33 in.) 
Weight  1.9 kg (4.2 lb) without mounting hardware 
3.4 kg (7.5 lb) with Pipe Mounting Hardware 
 
250 
14      19       24       29       34      39       44       49     54   (HART) 
13.5  18.5   23.5    28.5    33.5   38.5    43.5   48.5    53.5 (Analog)    
2100 
1800 
1500 
1200 
  900 
  600 
  300 
      0 
LOAD RESISTANCE  
2000 
HART  
COMMUNICATIONS 
RANGE 
Figure 27 – Supply Voltage Requirements, HART and Analog Versions 
Supply Voltage VDC