ABB TB82PH ATEX Product Instruction Manual
TB82PH-ATEX-EN-A 17
SECURITY MODE
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
Set Level
Calibrate
Output
Configure
Password
Enter 3 digit code
SECONDARY DISPLAY MODE
MAINTENANCE
Preventative Maintenance Tasks Interval (months)
Check, tighten and clean all wiring and connections 12
Clean and lubricate all gaskets and o-rings Each time seals are disturbed
Analyzer Validation Per user’s requirements (minimum 12)
Clean and inspect sensor As required by application
(minimum monthly)
Sensor calibration As required by application
(minimum monthly)
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
Temperature °C
Temperature °F
Current Output
Diagnostic Alarms
Reference Impedance
Millivolt Input
Software Revision
Back to Temp °C
Note.
If an ‘M’ appears after the
temperature display (e.g. 25°C
M
)
the TB82PH is programmed for
manual temperature compensation
(default 25°C).