ABB  TB82PH ATEX Product Instruction Manual 
TB82PH-ATEX-EN-A  17 
SECURITY MODE 
 
MEASURE         
 
CALIBRATE        
 
OUT/HOLD       
 
CONFIGURE        
 
SECURITY       
   
Set Level     
       Calibrate 
       Output 
       Configure 
   Password     
      Enter 3 digit code 
 
 
SECONDARY DISPLAY MODE 
 
 
 
MAINTENANCE 
 
Preventative Maintenance Tasks  Interval (months) 
Check, tighten and clean all wiring and connections  12 
Clean and lubricate all gaskets and o-rings  Each time seals are disturbed 
Analyzer Validation  Per user’s requirements (minimum 12) 
Clean and inspect sensor   As required by application 
(minimum monthly) 
Sensor calibration   As required by application 
(minimum monthly) 
 
MEASURE
CALIBRATE
OUT/HOLD
CONFIGURE
SECURITY
DISPLAY
Temperature °C
Temperature °F
Current Output
Diagnostic Alarms
Reference Impedance
Millivolt Input
Software Revision
Back to Temp °C
Note. 
If an ‘M’ appears after the 
temperature display (e.g. 25°C
M
) 
the TB82PH is programmed for 
manual temperature compensation 
(default 25°C).