TB82PH ATEX Product Instruction Manual ABB
14 TB82PH-ATEX-EN-A
Note.
Diagnostics is a function valid only with a TBX5
Advantage pH sensors or an AP200 with a
solution ground rod that is connected to Terminal
TB2-4 (green, SOL GND). Diagnostics must be
turned ‘OFF’ if TB5, AP100, AP300, 7650/7660,
ORP or sensors without an integral solution
ground are used.
Caution.
If a programming change has been made,
pressing the Exit to MEASURE key will activate
the SAVE? display (Figure 25). YES and NO
keys will also appear. Press the YES key to save
the changes. If the NO key is pressed any
configuration or programming changes will not be
saved.