Infecon. Follow proper infecon control procedures and inform paents to avoid showering and touching
the bandages.
The detectors and metal screening devices. Certain types of anthe devices, such as those used at
entrances or exits of department stores, libraries, and other public establishments, and airport security
screening devices may aect smulaon. Paents who are implanted with nonadjacent mulple leads and
paents who are sensive to low smulaon thresholds may experience a momentary increase in their
perceived smulaon, which has been described by some paents as uncomfortable or jolng. Paents
should use cauon when approaching such a device and should request assistance to bypass the device. If
they must proceed through the device, paents should turn o the EPG and proceed with cauon, being
sure to move through the detector quickly.
Implantaon of mulple leads. If mulple leads are implanted, leads and extensions should be routed in
close proximity. Nonadjacent leads can possibly create a conduit for stray electromagnec energy that could
cause the paent unwanted smulaon.
Handling and Implementaon
Single use only. The EPG header is intended for single use only.
Exposure to liquids. Device components should be stored where they will not be exposed to liquids or
excessive moisture, which can damage the package materials and components.
Care and handling of components. Use extreme care when handling system components prior to use.
Excessive heat, tracon, bending, and twisng, or the use of sharp instruments may damage components
and cause them to fail.
Package or component damage. Do not use the device if the package or components show signs of
damage. Return any suspect components to
Abbo
Medical
for evaluaon.