Connecng the Mullead Trial Cable to the EPG
To connect a mullead trial cable to the EPG:
NOTE: For further instrucons, refer to the clinician’s manual for the mullead trial cable.
1. Remove the EPG header cap to expose the EPG smulator port.1.
2. Carefully slide the connector end of the mullead trial cable into the EPG smulator port.2.
CAUTION:
▪
The EPG is not sterile. When connecng and tesng the EPG with the mullead trial cable, keep the
▪
EPG outside of the sterile eld.
▪ To avoid delivering jolng smulaon to the paent, turn o smulaon before plugging in the▪
mullead trial cable.