▪ Smulaon will automacally turn o when the baeries cannot support smulaon. Replace the EPG’s▪
baeries to communicate with the EPG, turn smulaon on, change smulaon sengs, or retrieve
smulaon usage informaon.
NOTE: If the baeries deplete during a trial, instruct the paent to refrain from replacing the
baeries to avoid disturbing the bandaging and lead posion.
Seng Up Communicaon Between the EPG and the Clinician
Programmer and Paent Controller
Before starng the surgical procedure, set up communicaon between the EPG and the clinician
programmer and paent controller. The following steps outline the procedure for seng up communicaon.
For further instrucons, refer to the clinician’s programming manual.
1. Place the magnet perpendicular to and centered directly over the front of the EPG for 10 seconds.1.
The indicator light on the EPG will start ashing.
2.
Set up communicaon using the clinician programmer app or the paent controller app following the
2.
direcons shown on the screen.
Aer seng up communicaon, the EPG is paired with that clinician programmer and paent controller. You
will not need to pair them again unless another paent controller or more than four clinician programmers
are paired to the EPG.
NOTE: Store paired EPGs and paent controllers together to reduce the number of mes you need to set
up communicaon.