Smulaon will automacally turn o when the baeries cannot support smulaon. Replace the EPG’s
baeries to communicate with the EPG, turn smulaon on, change smulaon sengs, or retrieve
smulaon usage informaon.
NOTE: If the baeries deplete during a trial, instruct the paent to refrain from replacing the
baeries to avoid disturbing the bandaging and lead posion.
Seng Up Communicaon Between the EPG and the Clinician
Programmer and Paent Controller
Before starng the surgical procedure, set up communicaon between the EPG and the clinician
programmer and paent controller. The following steps outline the procedure for seng up communicaon.
For further instrucons, refer to the clinician’s programming manual.
1. Place the magnet perpendicular to and centered directly over the front of the EPG for 10 seconds.1.
The indicator light on the EPG will start ashing.
2.
Set up communicaon using the clinician programmer app or the paent controller app following the
2.
direcons shown on the screen.
Aer seng up communicaon, the EPG is paired with that clinician programmer and paent controller. You
will not need to pair them again unless another paent controller or more than four clinician programmers
are paired to the EPG.
NOTE: Store paired EPGs and paent controllers together to reduce the number of mes you need to set
up communicaon.