1. If any of the lead or extension contacts came in contact with body uid or saline, thoroughly clean the1.
contacts with sterile deionized water or sterile water for irrigaon and dry them completely.
CAUTION: Do not bend the lead sharply or it may be damaged.
NOTE: The EPG header comes with open ports and aached port plugs. Do not remove the port plugs
from the header. If only using one lead or extension, use the lower port labeled 18 and insert a port
plug in the unused port.
2. Push and hold the buon on the EPG header to open the EPG header ports.2.
3. Using clean gloves, carefully slide the lead or extension into the EPG header unl you reach a hard stop3.
and all the contact bands are fully inside the connector assembly and hidden from view.
25