Important Control Solution Information
Use only AlphaTRAK™ Control Solution with the AlphaTRAK •
Meter.
Gently invert the AlphaTRAK Control Solution bottle two to •
three times before each use.
To avoid contamination of AlphaTRAK Control Solution, do not •
touch the tip of container to any surface and discard the first two
drops.
Replace the cap on the control solution bottle immediately after •
use.
Do not use control solution past the expiration date.•
Do not add water or any liquid to control solution.•
Control solution tests are specified to be accurate only when •
used between 59° – 104° F (15° 40° C).
Caution: Results from AlphaTRAK Control Solution tests do
not reflect the animals blood glucose level.
The control solution test results should fall within the •
appropriate range printed on the test strip vial.
Important: The control solution range is a target range for control
solution only. It is not a target range for the animals blood glucose
level.
If control solution results are out of range, repeat the test.•
If the control solution test results continue to fall outside the •
appropriate range, do not use the AlphaTRAK Meter to test
the animal’s blood glucose. Contact Customer Care (USA and
Canada) call (888) 299-7416 or Technical Enquiries (UK) call
+ 44 (0) 1628 644354.
For more information on how to obtain AlphaTRAK Control •
Solution, contact a veterinarian/veterinary surgeon, Customer
Care (USA and Canada) or Technical Enquiries (UK).
The AlphaTRAK™ Control Solution
AlphaTRAK Control Solution is a red liquid that
contains a fixed amount of glucose that is used to:
Practice testing without having to use animal blood.•
Ensure that the meter and test strips are working •
together properly.
A control solution test should be performed:
After setting up the AlphaTRAK Meter and before •
the first blood test.
When a new vial of test strips is opened.•
When blood sample results are questionable and confirmation •
that the monitor and test strips are working together properly is
required.
If the test strip vial was not closed properly.•
If the meter is dropped.•
Upon the advice of a veterinary professional.•
AlphaTRAK Control Solution is good for three months after
opening the bottle or until the expiration date printed on the label,
whichever comes first.
Important: Count forward three months from the date a new bottle
of control solution is opened. This is the discard date. Write this date
on the side of the control solution bottle.