Ventricular Sensing
In CRTPs, Ventricular Sensivity should be programmed to the highest seng (lowest sensivity) that will
provide ventricular sensing with adequate sensing margin. Le ventricular lead dislodgement to a
posion near the atria can result in atrial oversensing and ventricular inhibion.
Perform a thorough assessment of device funcon following exposure to any of the above.
Magnec Resonance Imaging (MRI)
MR Condional pulse generators and CRTPs. Tesng has demonstrated that the Abbo Medical
MR Condional system is condionally safe for use in the MRI environment when used according to the
instrucons in the MRIReady Systems manual. The Abbo Medical MR Condional system includes an
Abbo Medical MR Condional device connected to one or more Abbo Medical MR Condional leads.
MRI untested pulse generators and CRTPs. “Untested” indicates that the device has not been tested, and
its use in an MRI environment is not determined.
Scanning paents who have other MR Condional devices that are not implanted in cardiac ssue is
acceptable provided all MR Condional requirements for each implanted device are met.
Backup VVI
In rare instances, the device may revert to Backup VVI operaon at the sengs listed in the table below.
These values are not programmable.
When the device has reverted to Backup VVI operaon, the programmer displays a popup message
indicang that the device is operang at the Backup VVI values. Press the Connue buon and follow the
onscreen instrucons.