Ventricular Sensing
▪
In CRT‑Ps, Ventricular Sensivity should be programmed to the highest seng (lowest sensivity) that will
▪
provide ventricular sensing with adequate sensing margin. Le ventricular lead dislodgement to a
posion near the atria can result in atrial oversensing and ventricular inhibion.
▪ Perform a thorough assessment of device funcon following exposure to any of the above.▪
Magnec Resonance Imaging (MRI)
▪
MR Condional pulse generators and CRT‑Ps. Tesng has demonstrated that the Abbo Medical
▪
MR Condional system is condionally safe for use in the MRI environment when used according to the
instrucons in the MRI‑Ready Systems manual. The Abbo Medical MR Condional system includes an
Abbo Medical MR Condional device connected to one or more Abbo Medical MR Condional leads.
▪ MRI untested pulse generators and CRT‑Ps. “Untested” indicates that the device has not been tested, and
▪
its use in an MRI environment is not determined.
▪ Scanning paents who have other MR Condional devices that are not implanted in cardiac ssue is▪
acceptable provided all MR Condional requirements for each implanted device are met.
Backup VVI
▪
In rare instances, the device may revert to Backup VVI operaon at the sengs listed in the table below.
▪
These values are not programmable.
▪
When the device has reverted to Backup VVI operaon, the programmer displays a pop‑up message
▪
indicang that the device is operang at the Backup VVI values. Press the Connue buon and follow the
on‑screen instrucons.