remote possibility for a device to intermiently oversense while the CT scanning beam is directly over the
implanted device.
Sterilizaon
The package contents have been sterilized with ethylene oxide before shipment. This device is FOR
SINGLE USE ONLY and is not intended to be resterilized.
Do not implant the device when the sterility indicator within the inner package is purple, because it might
not have been sterilized, or when the storage package has been pierced or altered, because this could
have rendered it nonsterile.
Storage and Handling
Store the device at temperatures between 15°C (59°F) and 30°C (86°F). During transportaon and
handling the device can be exposed to temperature excursions between 20°C and 60°C (-4°F and 140°F).
Store the device in a clean area, away from magnets, kits containing magnets, and sources of
electromagnec interference to avoid device damage.
Do not implant a device that has been dropped on a hard surface while outside of its intact shelf package,
or from a height of more than 61 cm (24 in.) while within its intact shelf package. Sterility, integrity, or
funcon cannot be guaranteed under these condions, and the device should be returned to
Abbo
Medical
for inspecon.
Aer cold storage, allow the device to reach room temperature before programming or implanng the
device because cold temperature may aect inial device funcon.