As required by EN45502‑2‑1 Clause 27.4, the device interference mode of operaon is characterized as
follows:
▪ The atrial noise mode is “pacing o” for EMI frequencies below approximately 30 Hz and “xed rate▪
pacing” for frequencies above approximately 30 Hz.
▪
The ventricular noise mode is “xed rate pacing” for EMI frequencies of 16.6 Hz‑167 kHz.
▪
Temperature Eects
Pacing parameters such as Pulse Rate, Pulse Width, Pulse Amplitude and Sensivity meet the nominal
tolerances dened in the programmer’s on‑screen help over the temperature range of 25°C to 45°C (±2°C).
Input Impedance
Table 13. Input impedance
Measurement
Range
Input impedance 30‑75 kΩ
Eecve Pacing Capacitance
Table 14. Eecve pacing capacitance
Measurement
Range
Eecve pacing capacitance 4.7 µF
±10%