Test Pulse Sensivity
Sensivity was measured using the test pulse shown in the gure below.
Table 15. Test pulse sensivity (mV)—Posive signals, ventricular channel, 37°C
Programmed nominal Min Typical Max
0.5 0.35 0.40 0.65
1.0 0.7 1.0 1.3
7.0 4.9 7.3 9.1
12.5 8.75 12.9 16.25
Table 16. Test pulse sensivity (mV)—Negave signals, ventricular channel, 37°C
Programmed nominal Min Typical Max
-0.5 -0.35 -0.40 -0.65
-1.0 -0.7 -1.0 -1.3
-7.0 -4.9 -7.3 -9.1
-12.5 -8.75 -12.9 -16.25