EasyManua.ls Logo

Abbott Endurity Core - Environmental and Medical Therapy Hazards

Abbott Endurity Core
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Therapeuc Radiaon
See Warnings (page 9)
Hospital and Medical Environments
High Radiaon Sources
Do not direct high radiaon sources such as cobalt 60 or gamma radiaon at the pulse generator. If a
paent requires radiaon therapy in the vicinity of the pulse generator, place lead shielding over the
device to prevent radiaon damage, and conrm the device is funconing aer treatment.
Medical Lithotripsy
See Warnings (page 9)
Environmental and Medical Therapy Hazards
Advise paents to avoid devices that generate a strong electric or magnec interference (EMI). EMI could
cause malfuncon of or damage to the pulse generator, resulng in nondetecon or delivery of
unneeded therapy. Moving away from the source or turning it o will usually allow the pulse generator to
return to its normal mode of operaon.
Advise paents to avoid any areas marked with a “No Pacer” symbol.
17

Related product manuals