Therapeuc Radiaon
▪
See Warnings (page 9)
▪
Hospital and Medical Environments
High Radiaon Sources
▪ Do not direct high radiaon sources such as cobalt 60 or gamma radiaon at the pulse generator. If a▪
paent requires radiaon therapy in the vicinity of the pulse generator, place lead shielding over the
device to prevent radiaon damage, and conrm the device is funconing aer treatment.
Medical Lithotripsy
▪
See Warnings (page 9)
▪
Environmental and Medical Therapy Hazards
▪
Advise paents to avoid devices that generate a strong electric or magnec interference (EMI). EMI could
▪
cause malfuncon of or damage to the pulse generator, resulng in non‑detecon or delivery of
unneeded therapy. Moving away from the source or turning it o will usually allow the pulse generator to
return to its normal mode of operaon.
▪
Advise paents to avoid any areas marked with a “No Pacer” symbol.
▪
17