NOTES:
▪ When connecng leads to the pulse generator, make sure that you plug the correct lead into the correct▪
lead receptacle. For sensing and pacing, this is important to ensure that atrial and ventricular signals are
correctly recorded and that pacing pulses are delivered in the desired chamber.
▪ The IS4‑LLLL lead receptacle can only be used with IS4‑LLLL le heart leads.▪
Programming Guidelines
For a list of all programmable parameters and sengs, refer to the programmer's on‑screen help.
Removing a Magnet
To interrogate the device, remove the magnet from the programmer telemetry wand. A magnet will interfere
with proper telemetry.
Temporary Programming
These devices feature Temporary Programming to aid the clinician in diagnosing and treang the paent. The
clinician can temporarily program parameters to assess their eects with the ability to quickly cancel or
permanently program the seng. For more informaon, refer to the programmer's on‑screen help.
Preset Programmed Sengs
Shipped Sengs
The device's parameter sengs are preset when the device is manufactured. For addional informaon, refer
to the programmer's on‑screen help.
29