MRI Safety Informaon
MR Condional pulse generators and CRT‑Ps are condionally safe for use in the MRI environment when used
in a complete MR Condional system and according to the instrucons in the MRI‑Ready Systems manual.
Scanning under dierent condions may result in severe paent injury, death, or device malfuncon.
Summary of Safety and Clinical Performance
A summary of the safety and clinical performance for this device is available at
hps://ec.europa.eu/tools/eudamed. Search for the device using the Global Trade Item Number (GTIN; also
known as UDI) provided on the Informaon screen of the clinician programmer applicaon. This is the SSCP
locaon aer the launch of the European Database on
Medical Devices/Eudamed.
Contraindicaons
Implanted Cardioverter‑Debrillator (ICD). Single‑chamber pulse generators, dual‑chamber pulse generators,
and CRT‑Ps are contraindicated in paents with an implanted cardioverter‑debrillator.
Single‑Chamber Ventricular Demand Pacing is relavely contraindicated in paents who have demonstrated
pulse generator syndrome, have retrograde VA conducon, or suer a drop in arterial blood pressure with the
onset of ventricular pacing.
Data Security
Abbo
Medical
takes a broad and deep approach to ensuring the safety, security, and privacy of the paent
informaon and data on our devices and systems connecng paents to healthcare providers and clinics.
Paents, clinical sta, and hospital IT sta do not require any special training, and do not need to congure
7