AF Suppression (dualchamber pulse generators, CRTPs) smulaon is not recommended in paents who
cannot tolerate high atrialrate smulaon.
DualChamber Pacing (dualchamber pulse generators, CRTPs), though not contraindicated for paents with
chronic atrial uer, chronic atrial brillaon, or silent atria, may provide no benet beyond that of single-
chamber pacing in such paents.
SingleChamber Atrial Pacing is relavely contraindicated in paents who have demonstrated compromise of
AV conducon.
Device Communicaon
Communicaon with the device can be aected by electrical interference and strong magnec elds. If
this is a problem, turn o nearby electrical equipment or move it away from the paent and the
programmer. If the problem persists, contact Abbo
Medical
.
Subopmal RF Communicaon
The
Merlin
™ PCS or
Merlin
™ 2 PCS indicate the quality of the RF communicaon by the telemetry
strength indicator LEDs on both the programmer and the
Merlin
™ Antenna. Below is a list of potenal
causes to subopmal radio communicaon.
Table 4. Possible causes and soluons for subopmal RF communicaon
Possible Causes Soluons
The Merlin Antenna orientaon/
locaon is subopmal.
Move or reorient the Merlin Antenna slightly. Make sure that the front of the
Merlin Antenna faces the implantable device.