Table 8. Expected device lifeme
Model name Model number Device type Expected device lifeme
Assurity MRI PM2272 DR 9.9 years
Zenex MRI PM2282 DR 9.9 years
Allure RF PM3222 CRTP 8.3 years
Quadra Allure MP RF PM3262 CRTP 7 years
Quadra Allure PM3542 CRTP 8.6 years
Quadra Allure MP PM3562 CRTP 7.3 years
NOTE: For SR models, VVI is the most comprehensive pacing mode.
NOTE: For DR models, DDD is the most comprehensive pacing mode.
NOTE: For CRTP models, DDD BiV is the most comprehensive pacing mode.
Elecve Replacement Indicator (ERI)
ERI (or Recommended Replacement Time) is the point at which baery voltage has dropped to the lowest
capacity that will maintain adequate pulse generator operaon for a nominal period of before end of life
(EOL).
The me between ERI and EOL for all devices is 6 months.
When the device reaches ERI, a number of indicators alert the clinician to this condion. For informaon on
these condions, refer to the programmer's onscreen help.