"E-4" Result
What It Means:
There may be a problem with the test strip.
What to Do:
Monitor your blood β-Ketone again with a
new test strip. If E-4 shows on the display
window again, contact your healthcare
professional immediately.
Important:
It is recommended that you repeat the blood β-Ketone
test with a new test strip when:
HI appears in the display window.
Your result is unusually high.
You question your result.
You obtain a 0.0 mmol/L blood β-Ketone result BUT your
blood glucose is higher than 16.7 mmol/L (300 mg/dL).
Important: A result that is incorrect may have a serious
medical outcome. Consult your healthcare professional
before changing your diabetes medication programme.
35 36
CHAPTER 6
Doing A Control
Solution Test
Why Do A Control Solution Test?
The control solution test tells you that your meter and test
strips are working correctly. A control solution test is similar
to when you monitor your blood
glucose or blood β-Ketone, except you
use an Optium or MediSense Control
Solution. You do not use a drop of
blood in a control solution test.
When Is Doing A Control Solution
Test Recommended?
When you question your results.
To make sure that your meter and
test strips are working properly.
Important Information about
Control Solution Testing
For more detailed information on how to obtain control
solutions, please contact Customer Service in the UK on
0500 467 466, or in Ireland on 1800 776633.
For more detailed information about control solutions, read
the control solution instructions for use.
Optium or MediSense Control Solutions may be used for
glucose or β-Ketone control solution testing.
Do not use the control solution if the expiry date has passed.
Check the expiry date printed on the control solution bottle.
When you open a control solution bottle for the first time,
count forward 90 days and write this date on the control
solution bottle using a pen that won’t smear or wipe off.
Throw away any remaining solution after this date.
Control
Solution Testing
Questions? Call
Customer Support:
UK: 0500 467 466
Ireland: 1800 776633
Monitoring Blood
β-Ketone