58 Technical Specications
Technical Features
Maximum Infusion
Pressure Generated by
Pump
200 kPa
Maximum Time for
Activation of the Occlusion
Alarm
Note: Changes in temperature and infusion set length
may affect the time to detect an occlusion.
The time to detect an occlusion is 5 hours or less when
infusing under the following conditions:
• basal delivery ow rate of 0.15 mL/hr
• infusion set length of 60 cm
• occlusion introduced at patient end of infusion set
• ambient temperature of 68 ± 4 °F (20 ± 2 °C)
• ambient humidity of 65 ± 5 % RH
The time to detect an occlusion is 2 hours or less when
infusing under the following conditions:
• basal delivery ow rate of 0.70 mL/hr
• infusion set length of 60 cm
• occlusion introduced at patient end of infusion set
• ambient temperature of 68 ± 4 °F (20 ± 2 °C)
• ambient humidity of 65 ± 5 % RH
Maximum Volume of
Unintended Bolus After
Occlusion
The pump, when pumping at 0.7 mL/hr, has a maximum
unintended bolus volume of 0.8 mL under the following
conditions:
• infusion set length of 60 cm
• occlusion introduced at patient end of infusion set
• ambient temperature of 68 ± 4 °F (20 ± 2 °C)
System Operating
Conditions
To maintain all essential performance requirements,
the operating temperature range of the pump, battery
charging system, and carrying accessory is between
41 °F (5 °C) and 104 °F (40 °C), inclusive, within
the humidity range of 15% to 90% non-condensing,
inclusive, and within the atmospheric pressure range of
70 kPa to 106 kPa, inclusive.
Pump Kit Storage and
Transport Conditions
(including the Pump,
Battery Charging System,
and Carrying Accessory)
-4 °F to 41 °F (-20 °C to 5 °C) with uncontrolled humidity
41 °F to 104 °F (5 °C to 40 °C), up to 90% relative
humidity non-condensing
104 °F to 140 °F (40 °C to 60 °C), up to 15% relative
humidity non-condensing
Note: At atmospheric pressure ranging from 53.3 kPa to
106 kPa, inclusive.