17Initial Setup
Set Extra Dose
2.5 Set Extra Dose
1. Set up extra dose.
• Press NO to skip extra dose set up
and disable extra dose.
• Press YES to set up extra dose
parameters and enable extra dose.
2. Set the extra dose volume.
a. Use the arrow keys to set the
value.
b. Press NEXT once extra dose is set
to the desired value.
3. Set the lock-out time.
a. Use the arrow keys to set the
value.
b. Press NEXT once lock-out time is
set to the desired value.
Note: The minimum lock-out time is
1 hour.
The delivery of an extra dose can be enabled. The extra dose volume
range is 0.1 mL to 0.3 mL and can be adjusted in increments of 0.05 mL.
Note: If the patient is not going to receive an extra dose, when prompted
to "Set up Extra Dose?", select "NO". If you accidentally select "YES"
and do not want to set an extra dose, set the extra dose value to 0 mL.
A lock-out time interval for extra dose deliveries must be congured.
The lock-out time is the interval from the end of the delivery of the most
recent extra dose or loading dose to the next extra dose becoming
available. During the extra dose lock-out time, no extra dose deliveries
can be performed by the pump. The lock-out time prevents the delivery
of an unlimited number of extra doses. The lock-out time range is 1 hour
to 24 hours (in 15-minute increments).
Note: When a loading dose is administered, the extra dose lock-out timer
will be restarted or "re-set.”