57Technical Specications
Technical Features
11. Technical Specications
Pump Dimensions 170 x 76 x 33 mm (6.7 x 3.0 x 1.3 in)
Pump Weight 285 g (10.0 oz.), including battery
Battery 3.6 V Li-Ion / 2.35 Ah / 8.46 Wh
Pump Ingress Protection
Rating
IP22
Pump Security Lock
Levels
Protected Clinician Mode
Permissible Orientations
of the Pump
No specic orientation required
Continuous Dose Delivery
Flow Rates
Programmable from 0.15 mL/hr to 1.25 mL/hr with
0.01 mL/hr increments
Number of Selectable Flow
Rates
Up to 3, based on conguration
Average Flow Rate During
Extra Dose and Loading
Dose Delivery
5.5 mL/hr
Priming Volume From 0.15 mL to 0.6 mL
Extra Dose Volume Programmable from 0.1 mL to 0.3 mL in 0.05 mL
increments
Lock-out Time Interval
Between Extra Doses
Programmable from 1 hour to 24 hours in 15 minute
increments
Loading Dose Volume Programmable from 0.1 mL to 3.0 mL in 0.1 mL
increments
Lock-out Time Interval
Between Loading Doses
Programmable from 3 hours to 8 hours in 1 hour
increments
Typical Service Life of the
Pump
The pump is expected to have a service life of 3 years.
Time to Bring System to
Operating Temperature
The operating temperature of the system is between
41 °F (5 °C) and 104 °F (40 °C). It may, however, be
stored between -4 °F (-20°C) and 140 °F (60 °C).
If the system is stored or transported at the maximum or
minimum temperatures allowed for storage and transport,
it requires 30 minutes in a 68 °F (20 °C), environment to
reach operating temperature.
11.1 Technical Features