23Change Therapy Settings
Access Clinician Settings
3. Change Therapy Settings
If changes to patient therapy or other pump settings are needed after the
initial setup is complete, you can access the clinician settings using the
following procedure. The clinician settings will allow you to:
Set continuous infusion rate parameters
Set extra dose parameters (or set to 0 to disable)
Set loading dose parameters (or set to 0 to disable)
Change PIN
Set pump language
1. Navigate to Clinician Settings.
a. Press MENU to display pump
Menu Options.
b. Use the arrow keys to navigate to
Clinician Settings.
c. Press SELECT.
3.1 Access Clinician Settings