103
1. With the meter off, wipe the
test strip slot with a cotton
swab lightly moistened with
water.
2. Open the drum door: Hold the
meter in your right hand and
place the thumb of your left
hand into the recess below the
drum door. Then pull the drum
door with your ngers slightly
to the front and lift it up.
3. Wipe the measuring window
and the area surrounding it
with a cotton swab lightly
moistened with water.
4. Dry the meter with a soft cloth
or gauze and make sure that
no solution is seen in any
openings.
5. Remove any uff or lint that
may remain.
6. Close the drum door until it
snaps.
7. Perform a control test (see
Chapter 2, “Performing a
Control Test.”).
man_07699336001_01_EN.indb 103 28.09.2015 13:27:09