EasyManuals Logo

Accu-Chek Compact Plus Owner's Booklet

Accu-Chek Compact Plus
244 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #104 background imageLoading...
Page #104 background image
104
Not enough blood or control
solution was drawn into the
test strip for measurement.
Discard the test strip and
repeat the test.
Blood or control solution was
applied to the test strip before
the ashing test strip and drop
symbol appeared on the
display. – Discard the test strip
and repeat the test.
The test strip was bent before
the test started or during the
test. – Discard the test strip
and repeat the test.
A test strip is stuck in the
meter. – Hold the meter
upright so the test strip is
pointing downwards. Slide and
release
to turn off the
meter and release the test
strip. If the test strip does not
drop out, remove it manually,
even if you feel some
resistance.
The test strip may be
damaged. – Discard the test
strip and repeat the test.
The test strip was pulled out
of the meter. – Slide and
release
to turn off the
meter. Only remove the test
strip from the meter by sliding
and releasing
.
Condensation has affected the
electronics. – Allow the meter
time to dry slowly.
man_07699336001_01_EN.indb 104 28.09.2015 13:27:09

Table of Contents

Other manuals for Accu-Chek Compact Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Compact Plus and is the answer not in the manual?

Accu-Chek Compact Plus Specifications

General IconGeneral
Measurement Range10-600 mg/dL
Sample Size1.5 µL
Test Time5 seconds
Automatic CodingYes
DisplayLCD
Measurement MethodElectrochemical
Battery Type2 x AAA
Battery LifeApprox. 1000 tests
Data TransferInfrared (IR)
Storage Conditions-25°C to +70°C
Operating Conditions10-40°C, 10-90% RH