EasyManuals Logo

Accu-Chek Compact Plus Owner's Booklet

Accu-Chek Compact Plus
244 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #106 background imageLoading...
Page #106 background image
106
The test strip is stuck in the test
strip slot or the test drum does
not rotate. Slide and release
to turn off the meter. Open
the drum door and, if applicable,
pull the test strip out of the
meter and discard it. Press the
red release button. Take the test
drum out and reinsert it. The next
time you slide and release
to turn on the meter and perform
a test, the drum symbol is
displayed and the bottle symbol
ashes. The 90-day use-by
period is no longer calculated for
this test drum.
The temperature is below or
above the range for performing a
test. – Slide and release
to
turn off the meter. Move to an
area where the temperature is
between 54 and 104 °F and wait
for the temperature of the meter
to adjust to this temperature. Do
not articially heat or cool the
meter.
Not enough blood or control
solution was drawn into the test
strip for measurement. – Discard
the test strip and repeat the test.
man_07699336001_01_EN.indb 106 28.09.2015 13:27:10

Table of Contents

Other manuals for Accu-Chek Compact Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Compact Plus and is the answer not in the manual?

Accu-Chek Compact Plus Specifications

General IconGeneral
Measurement Range10-600 mg/dL
Sample Size1.5 µL
Test Time5 seconds
Automatic CodingYes
DisplayLCD
Measurement MethodElectrochemical
Battery Type2 x AAA
Battery LifeApprox. 1000 tests
Data TransferInfrared (IR)
Storage Conditions-25°C to +70°C
Operating Conditions10-40°C, 10-90% RH

Related product manuals