4. The time and date appear on
the display with a ashing test
strip and drop symbol.
5. Select the control solution you
want to test.
6. Remove the control bottle cap
and wipe the tip of the bottle
with a tissue.
7. Invert the bottle and hold it
downwards at an angle.
Squeeze it gently until a small
drop forms at the tip.
8. Apply the drop to a clean and
easily washable surface made
of glass, porcelain or stainless
steel, such as a plate or spoon.
NOTE
Control solution you apply must come from a drop that has formed at the bottle tip. Do not use the drop
if it has run down the outside of the tip. This may lead to results that are too high.
Control solution must have been freshly applied to the surface. The control solution must not be left
exposed on the surface for any length of time, or it may give results that are too high.
Level 1 Level 2
OR
man_07699336001_01_EN.indb 37 28.09.2015 13:26:19

Table of Contents

Other manuals for Accu-Chek Compact Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Compact Plus and is the answer not in the manual?

Accu-Chek Compact Plus Specifications

General IconGeneral
Measurement Range10-600 mg/dL
Sample Size1.5 µL
Test Time5 seconds
Automatic CodingYes
DisplayLCD
Measurement MethodElectrochemical
Battery Type2 x AAA
Battery LifeApprox. 1000 tests
Data TransferInfrared (IR)
Storage Conditions-25°C to +70°C
Operating Conditions10-40°C, 10-90% RH

Related product manuals