Control Tests
4
Understanding Out‑of‑Range Control Results
WARNING
The control range applies only to a control result. It only indicates that the test strips and meter are
working properly. Do not use a control result to interpret blood glucose results.
If the control result is out of range, do not use the meter until you solve the problem. Check this list to
help solve the problem.
Troubleshooting Checks Action
Were the test strips or control solutions expired? Discard the test strips or control solution if either
is past the Use By date. If the control solution
was opened more than 3months ago, discard it.
Repeat the control test with an unexpired test
strip and an unexpired control solution.
Did you wipe the tip of the control solution bottle
before use?
Wipe the tip of the bottle with a tissue. Repeat
the control test with a new test strip and a fresh
drop of control solution.
Were the caps on the test strip container and the
control solution bottle always closed tightly?
Replace the test strips or control solution if you
think either was uncapped for some time. Repeat
the control test.
Was the test strip used immediately after it was
removed from the test strip container?
Repeat the control test with a new test strip and
a fresh drop of control solution.
Were the test strips and control solutions stored
in a cool, dry place?
Repeat the control test with a properly stored test
strip or control solution.
Did you follow the directions? Read the chapter Control Tests and repeat the
control test.
Did you choose the correct control solution level,
either 1 or 2, when you performed the test?
If you chose the wrong control solution level, you
can still compare the control result to the range
printed on the test strip container.
Are you still unsure of the problem? Contact the Accu‑Chek Customer Care Service
Center at 1‑800‑858‑8072.
54988_08100918001_EN.indb 42 9/30/16 5:04 PM

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Guide and is the answer not in the manual?

Accu-Chek Guide Specifications

General IconGeneral
Measurement Range10 to 600 mg/dL
Test TimeLess than 4 seconds
Battery LifeApproximately 1000 tests
ConnectivityBluetooth
Sample Size0.6 μL
WeightApprox. 40 g (1.4 oz) with battery
Battery Type2 x CR2032 lithium coin cells
DisplayLCD
Operating Conditions10-40°C (50-104°F), 10-90% relative humidity
Storage Conditions-25°C to +70°C (-13°F to +158°F)