6-4 After Data acquisition(Data acquisition)
・ When [Meas Time] is other than
***
not receiving the
E
command
: Data acquisition completed command
r
is issued to CPU.
・
In Interval/Repeat mode,
r
is issued every time of the measurement
・
On
E
command issued, the measurement is terminated even in the Repeat mode.
6226 PC
Data acquisition
completed command
r
Data Request command
D
Data Transfer
< Data format>
00/01/05□02:15:16□F□130dB+LF
00h00m00s+LF
Laeq:□130.0□□□La05:□130.0+LF
Lae□:□130.0□□□La10:□130.0+LF
□□□□□□□□□□□□□□La50:□130.0+LF
LMin:□130.0□□□La90:□130.0+LF
LMax:□130.0□□□La10:□130.0LF +LF
□ :Space (20)
+LF:Code linefeed(0A)
6-5 Lp-value acquisition
6226 PC
<Data format > (5Byte + CR )
130.0 + CR
6-6 Notes
・Flow control is not available in 6226.
・For communication command error, the
?
mark is returned.
・To return to Calibration, Stop measurement command
E
is used.
P :Lp acquisition
Data transfer
Data
Data