4 Connect the medical device
4.1 Connecting the medical device to the electrical network
Have your medical device connected to the mains power by an approved dental installation technician.
Switch the medical device OFF (position O) and check that the mains voltage is compatible with that indicated on the
medical device or its mains adapter. Next, connect the cord to the wall socket in compliance with the standards in force
in the country of use.
A different voltage would cause damage to the medical device and could injure the patient and the user. Any variation
in the electrical network voltage or electromagnetic field that is non-compliant with the limits in force, could interfere
with the medical device's operation.
Medical devices equipped with a protective earth must be connected to a supply network equipped with a protective
earth.
Do not plug the medical device into an extension lead and do not put the mains cord in a cable cover or cable
tray.
If when using the medical device, a power outage can create an unacceptable risk, the user and the installer
must ensure that the medical device is connected to an appropriate power source such as an uninterruptable
power supply.
4.2 Connecting the medical device to the electrical network
1. Set the medical device's mains switch to "O" OFF position.
2. Connect the mains cord to the control unit's mains connector.
3. Connect the power lead to the mains socket.
User Manual | Piezotome Cube | J50101 | V4 | (17) | 03/2018 | NO37EN010D - Page 9/39