5 Installing the medical device
Place the medical device in the position that is suitable for your activity.
The medical device must be placed on a secure and flat surface or a surface with a maximum slope of five degrees.
Check that the cords do not hinder the movement or free circulation of anyone.
Adjust the position of your medical device to correspond to your angle of vision and the characteristics of your
workstation, e.g. lighting or distance between the user and the medical device.
Ensure that your medical device is readily accessible. The disconnecting devices - the switch and the power plug - must
be easy to locate and access.
Do not install your medical device near or on another device.
5.1 Install cords
Never wrap the handpiece cord around the medical device.
Make sure that it is not possible to wheel over or walk on the different cords.
The cord attached to its handpiece must be easily accessible. Make sure that the cord is slack during use.
5.2 Installing the control pedal
Connect the footswitch cord to the back of the medical device.
The control pedal must be positioned near the feet of the operator and must be readily accessible.
5.3 Connecting the handpiece
Connect the cord and handpiece assembly to the connector on the control unit panel.
5.4 Installing an irrigation line
1. Remove the irrigation line from its sterilisation bag.
2. Remove the clips from their bag.
3. Open the cassette holder on the right side of the medical device.
4. Insert the cassette and close the holder.
5. Connect the end of the irrigation line, the long infusion line, to the handpiece
6. Working your hands along the handpiece cord, attach the irrigation line to the handpiece cord using the clips.
Any excess irrigation line will be at the medical device unit and will not hamper use of the handpiece.
7. On the short infusion line side, pierce the irrigation solution bag with the perforator.
8. When the medical device is switched ON, open the perforator cap and purge the irrigation system.
User Manual | Piezotome Cube | J50101 | V4 | (17) | 03/2018 | NO37EN010D - Page 11/39