2 Required information
2.1 Intended use
The intended use of the Piezotome Cube SATELEC is to provide utilities and to act as a base for various dental
instruments and accessories.This device is intended to be used by qualified professional dentists.
2.2 Indication for use
The Piezotome Cube is a command unit used in combination with the following items.
An intraoral surgery ultrasonic handpiece and an intraoral surgery tip. This combination is intended for use in intraoral
surgery procedures, including osteotomy osteoplasty, syndesmotomy, membrane detachment and bone expansion.
2.3 Operating principle
An electrical signal emitted by the medical device is supplied to the ultrasonic handpiece. This is connected to the
medical device via a cord. The handpiece comprises a piezoelectric ceramic transducer, which transforms the
electrical signal into ultrasonic vibrations.
Mechanical vibrations are transmitted to an intraoral surgery dental tip attached to the end of the ultrasonic
handpiece.
The medical device must be used with a Cube LED handpiece. Refer to the Cube LED [J28821] handpiece user manual
for further information.
2.4 Using accessories not supplied by the manufacturer
The handpiece is designed to operate with SATELEC, a company of Acteon group tips. The use of other makes of tips
will damage the handpiece and break the tips.
2.5 Connecting and disconnecting accessories during use
Do not tighten or loosen the tips when the handpiece is activated.
2.6 Repairing or modifying the medical device
Contact the supplier of your device. Using the services of an unapproved repairer could render your device dangerous
for you and your patients.
Do not repair or modify the device without seeking the prior permission of SATELEC, a company of Acteon group.
If the device is modified or repaired, specific checks and tests must be carried out to ensure that the medical device is
still safe to use.
In the event of doubt, contact an approved dealer or the SATELEC, a company of Acteon group Customer Service team:
www.acteongroup.com
satelec@acteongroup.com
SATELEC, a company of Acteon group, at the request of technical personnel working for the network of
approved dealers, will provide any information required to repair defective parts on which they may perform
repairs.
User Manual | Piezotome Cube | J50101 | V4 | (17) | 03/2018 | NO37EN010D - Page 5/39