8 Disinfection and sterilising
The instructions relating to cleaning, disinfection and sterilisation protocols for accessories supplied by SATELEC, a
company of Acteon group have been approved for each medical device and accessory. The applicable guides are listed
in the chapter Associated documentation page 3.
They can be downloaded at the following address: www.satelec.com/documents.
In all cases, the local regulations in force relating to the cleaning, disinfection and sterilisation protocols for
accessories take precedence over the information provided by SATELEC, a company of Acteon group.
8.1 Clean and disinfect the medical device
The medical device's control unit must be cleaned and disinfected daily.
The medical device's control pedal must be cleaned and disinfected daily.
The handpiece and its cord must be cleaned, disinfected and sterilised after use.
The autoclavable irrigation lines must be cleaned, disinfected and sterilised after each use.
The medical device must be in OFF or O stop position during cleaning and disinfecting procedures.
Refer to the instructions in the chapter Cleaning the irrigation system page 25.
Use alcohol disinfectant wipes.
Avoid using cleaning and disinfection products that contain flammable agents.
Otherwise, ensure that the product has completely evaporated or that there is no fuel left on the medical device and its
accessories before switching it on.
Do not use an abrasive product to clean the medical device.
Never apply sprays directly to the medical device to clean it. Always spray the product onto a wipe, then clean
the medical device.
The autoclavable irrigation lines must be cleaned, disinfected and sterilised after each use.
To prepare for cleaning, remove the various parts of the Piezotome Cube as shown here.