Pressing the foot-switch automatically activates the handpiece ultrasounds and the irrigation function.
The control footswitch fitted with its cord must be disconnected for daily cleaning using an alcohol disinfectant wipe.
The light function remains active for approx. 9 seconds after the pedal is released.
For further information, refer to the chapter entitled Pedal Overview page 1.
7.12 Mains Connector
The mains connector with its earthing pin is used to connect the device to the electrical network via a disconnectable
mains cord.
7.13 Switch
The mains switch is used to switch on (position I) or to stop (position O) the medical device.
7.14 Fuse recess
The recess holds two fuses designed to protect the medical device in the event of overvoltage or an internal fault.
Please read the instructions listed in the chapter Replacing the fuses page 25
7.15 Irrigation lines
The autoclavable irrigation lines can be reused after cleaning, disinfection and sterilisation.
After use, sterile irrigation lines should be discarded in a biomedical waste container.
Bottles or irrigation bags should not weigh more than one kilogram. A heavier container will make the medical device
tip over.
The medical device is not designed to deliver medicinal substances.
The medical device may only be used with bags or bottles of saline solution or sterile water.