6 Dispensing a treatment
6.1 Accessory usage conditions
The accessories of the Piezotome Cube must be cleaned, disinfected and sterilised before use.
Refer to the cleaning, disinfection and sterilisation protocols for accessories listed in the chapter Associated
documentation page 3.
This medical device is designed to be used in conjunction with a SATELEC, a company of Acteon group
handpiece and second-generation intraoral dental surgery tips.
6.2 Preparation for use
To prepare your medical device, follow the steps below:
1. Wear safety goggles and protective gloves.
2. Put the handpiece support in place.
3. Clean the unit with an alcohol disinfectant wipe.
1. Put the bracket in place.
2. Connect the handpiece cord to the connector on the front of the medical device.
3. Remove the handpiece support from its sterilisation pouch.
4. Remove the handpiece and its cord out of its sterilisation pouch.
5. Remove the wrench from its sterilisation bag.
6. Remove the tip from its sterilisation bag.
7. Screw the tip onto the handpiece, first manually and finishing with the wrench.
8. Place the handpiece on its support.
9. Put an irrigation solution bag in place on the bracket.
10. Remove the irrigation line and its cassette from its sterilisation bag or packaging if sterile.
11. Put in place the irrigation line and its cassette, up to the irrigation bag.
12. Switch on the medical device.
13. Check the irrigation settings according to the chosen tip and adjust the flow via the touch zones.
and .
14. Check the mode according to the chosen tip and adjust the active mode via the touch zones
and .
15. Above a water drain, check that the spray of the handpiece is working properly.
Your medical device is now ready to use.
6.3 Switching off the medical device
After installation and before first use, at the end of the day and following a period of prolonged non-use of the medical
device, it is important to clean the irrigation system.
This allows the autoclavable irrigation line to be cleaned prior to being disinfected, cleaned and sterilised.
When using irrigation solution bags to irrigate your medical device:
1. Disconnect the irrigation bag from the perforator of the irrigation line.
2. Dispose of the irrigation bag.
3. Dip the short end of the irrigation line in a recipient containing a hypochlorite solution diluted at less than 3%.
User Manual | Piezotome Cube | J50101 | V4 | (17) | 03/2018 | NO37EN010D - Page 15/39