4.
Press the Purge icon .
5. Operate the irrigation spray for two minutes to rinse the medical device's internal water system.
6. Refill the tank with demineralised or distilled water.
7. Rinse the irrigation system for two minutes.
When the irrigation system has been cleaned, perform the following operations:
1. Disconnect the handpiece and cord assembly and refer to the J12801 handpiece cleaning, disinfection and
sterilisation protocols.
2. Clean and disinfect the medical device as indicated in the chapter Clean and disinfect the medical device page 23.
3. Refer to the cleaning, disinfection and sterilisation protocols for SATELEC, a company of Acteon group
accessories listed in the chapter entitled chapter Associated documentation page 3.
Page 16/39 - User Manual | Piezotome Cube | J50101 | V4 | (17) | 03/2018 | NO37EN010D