Operator’s manual • X-Mind trium • V0G • (13) • 10/2017 • NTR0EN040G
A = X-ray emission switch
B = X-ray emission LED
• In case of CBCT scan, keep the switch pressed for the whole duration of the exam until the U-arm stops. If the switch is
released earlier, the X-ray emission and the carriage movements will be stopped and the exam is irremediably interrupted
(“dead man” exposure mode); in this case, the device must be reset and the patient positioning redone.
• In case the scout view has been selected, evaluate the scout view to establish whether the patient is positioned correctly.
• If the patient is positioned correctly, press YES to continue with the acquisition, otherwise, press NO to repeat the patient
positioning procedure.
CAUTION :
Monitor the patient throughout the entire exposure. In the event of an emergency, release the X-ray exposure switch to stop the
U-Arm rotation and X-ray emission.
In the event that the U-Arm rotation and/or X-ray emission do not stop, press the remote emergency switch.