Close temple rest on the patient forehead and gently move the patient’s head to the correct position with his/her mid-sagittal
plane corresponding to the mid-sagittal LASER.
6.3 EXECUTION OF THE CBCT EXAM
• Once the patient is correctly positioned, press “O position” key.
• The green status indicator will be ON and the U-arm will reach the start position.
• Ask the patient to stay still during the whole exposure.
• Go out from the room to start exam; do not forget to observe the applicable radiation safety procedure.
• Start the CBCT scan or acquire a scout view of the patient by pressing the X-ray emission switch on the X-MIND trium remote
control. During this stage, the device emits X-rays; the X-ray emission LED on the control panel and the X-ray exposure LED on
the X-MIND trium remote control will light up.