66
Operator’s manual • X-Mind trium • V0G • (13) • 10/2017 • NTR0EN040G
ENGLISH
DAP values for CBCT exams
Patient gender Patient size
Tube voltage
[kV]
Tube current
[mA]
Dental FOV 40x40
Standard Medium High
[mGycm^2] [mGycm^2] [mGycm^2]
CHILD
SMALL 80 6 140,4 168,5 210,6
MEDIUM 80 8 187,2 224,6 280,8
LARGE 80 10 234,0 280,8 350,9
WOMAN
SMALL 85 6 165,4 198,5 248,2
MEDIUM 85 8 220,6 264,7 330,9
LARGE 85 10 275,7 330,9 413,6
MAN
SMALL 90 6 190,5 228,6 285,7
MEDIUM 90 8 254,0 304,8 381,0
LARGE 90 10 317,5 381,0 476,2
Patient gender Patient size
Tube voltage
[kV]
Tube current
[mA]
Dental FOV 60x60
Standard Medium High
[mGycm^2] [mGycm^2] [mGycm^2]
CHILD
SMALL 80 6 359,6 431,5 539,3
MEDIUM 80 8 479,4 575,3 719,1
LARGE 80 10 599,3 719,1 898,9
WOMAN
SMALL 85 6 424,3 509,2 636,5
MEDIUM 85 8 565,7 678,9 848,6
LARGE 85 10 707,2 848,6 1060,8
MAN
SMALL 90 6 489,1 586,9 733,6
MEDIUM 90 8 652,1 782,5 978,1
LARGE 90 10 815,1 978,1 1222,7
Patient gender Patient size
Tube voltage
[kV]
Tube current
[mA]
Dental FOV 80x80 / 110x80
Standard Medium High
[mGycm^2] [mGycm^2] [mGycm^2]
CHILD
SMALL 80 6 414,0 496,8 621,0
MEDIUM 80 8 552,0 662,5 828,1
LARGE 80 10 690,1 828,1 1035,1
WOMAN
SMALL 85 6 489,1 586,9 733,7
MEDIUM 85 8 652,1 782,6 978,2
LARGE 85 10 815,2 978,2 1222,8