3.9.2 Configuration parameter integrity check
Check is carried out every operating cycle.
If a configuration parameter is found defective, the message ā€œER XXā€ will be reported on display for half a
second every two seconds, where XX is the code of the wrong parameter, according to the following table:
CODE Meaning Effect on indicator Resolution
ER 01 Error in the CALIBRATION
parameters
Relays are deenergized.
Error code is reported on serial line.
Instrument must be
recalibrated
ER 02 Error in the TARE
parameters
Relays are deenergized.
Error code is reported on serial line.
Redo tare
ER 03 Error in DIVIS e SENSIB
parameters
Relays are deenergized.
Error code is reported on serial line.
Reconfigure parameters
ER 04 Error in
AVER and DEC. PNT
No effect
Error code is reported on serial line.
Reconfigure parameters
ER 05 Error in LIM AZM -
MOTION - ZERO
TRACKING parameters
Relays are deenergized.
Error code is reported on serial line.
Reconfigure parameters
ER 06 Error in SER P - SER A-
BAUD - DLY SND - DLY
RCV - AD485 parameters
No effect Probably malfunctions on
serial line operation. The error code is
reported on the serial line
Reconfigure parameters
ER 07 Error in the RELAY 1
parameters
Relay 1 is deenergized.
Error code is reported on serial line.
Reconfigure parameters
ER 08 Error in the RELAY 2
parameters
Relays are deenergized.
Error code is reported on serial line.
Reconfigure parameters