List of Coils supported by A100E
Coil
Number
Function
0 Power On Flag reset request
1 GROSS (Data = $0000) or NET (Data = $FF00) set
2 ZERO request
3 TARE request
Status word coding
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
POF 0 S
Diagnostic error code
G/N 0 0 0 0 0 R2 R1
POF Power On Flag. Indicates a return of power supply to the transmitter.
Flag is set to 1 by A100E and can be resetetd to 0 by host by forcing
Coil #1
S if = 1 indicates configuration in progress
Diagnostic err code error code (see “self diagnostic” paragraph
G/N if = 1 indicates that the instrument is in NET mode
R2,R1 if = 1 indicates that the relevant relay is energized