MAINTENANCE
30
[1]
1 - Turn off the system;
2 - Close tap water valve;
3 - Allow a minute for water pressure to drop in
the tubing
[2]
Access the rear panel where the deionization
module is located:
1 - Insert Adrotool in filter connector slot to
disconnect Adrona deionization module;
2 - Pull down the tool.
[3]
Now you can remove the deionization module from
connector tubes and remove the deionization
module from the system
[4]
Unpack spare deionization module. Remove both
protective ¼” JG plugs from DI module.
Option 1.
You need to push the little ring down and you will
be able to take out the plug (see the principle in
page 7, section” Introduction to John Guest tubing
system”)